Mettler Sonicator 740 Ultrasound

Mettler Sonicator 740 Ultrasound

The Sonicator 740 is a 1 and 3 MHz therapeutic ultrasound unit which comes with a 5 cm² dual frequency applicator. It is light-weight and portable and features a universal applicator cable to make changing applicators quick and easy. There is an optional rechargeable lithium ion battery pack that truly allows you to take therapy on the road with you. Large soft-touch controls let you set-up treatments quickly and easily and the backlit display shows all treatment parameters and conditions.

General Specifications:

  • ETL and C-ETL Listed: Model ME 740 (9801427)
  • Input: 100-240 VAC ±10%
  • Frequency: 60Hz ±3 and 50Hz ±3
  • Input Current: less than 0.6 amps
  • Optional Battery: Rechargeable Smart Lithium Ion Battery Pack rated at 10.8Vand 4.8Ah
  • Patents: U.S. and foreign patents applied for and granted including U.S. Patent Numbers 4,966,131 and 5,095,890
  • Timer Accuracy: ±10% of the indicated time>/li>
  • Maximum treatment time: 29 minutes
  • Size: 13.5 in (L) x 9.5 in (W) x 7 in (H)
  • Weight: 3.75 pounds without battery
  • 4.5 pounds with battery
Ultrasonic Generator Specifications:
  • Frequency: 1.0 MHz ±10%; 3.2 MHz ±10%; 3.3 MHz ±10%
  • Modes: Continuous, Pulsed 10, 20 and 50% Duty cycle
  • Pulse repetition rate: 100 Hz ±5% (Pulse Mode)
  • Pulse duration: 1, 2 or 5 msec ±5%
  • Temporal peak/average intensity ratio: 10:1, 5:1 and 2:1 ±5%
  • Maximum output power: 22 W with a 10 cm² applicator (ME 7410); 11 W with 5 cm² applicator (ME 7413); 2.2 W with a 1 cm² applicator (ME 7431)
  • Maximum intensity: 2.2 W/cm²
  • Indication accuracy: ±20% (for any level above 10% of maximum)
Ultrasonic Applicator Specifications:
  • Piezoelectric discs: The output transducer utilizes a barium titanate disc with a specially coated face.


Warning: Federal law (USA) restricts this device to sale by or on the order of a licensed practitioner